Abstract:In the analysis for anions in environmental water with ion chromatography (IC), the drip washing liquid with 1.8 mmol/L Na2HCO3 and 1.7 mmol/L NaHCO3 is adopted. The regenerative liquid is 36 mmol/L H2SO4. The system is composed of series chromatography of PacAs-14 and conductivity detector. For 10 μL sample, within 8 min, seven kinds of anions can be detected. The working condition and the affecting factors of IC is discussed. Its relative standard deviation is between 1.9% and 6.0%, with a recovery above 97%. The method is practical, rapid and reliable. It can completely meet the demand of the analysis for anion in environmental water samples.
郝原芳, 吴英杰, 何炼. 离子色谱法测定环境水中阴离子[J]. 地质与资源, 2005, 14(2): 149-150,148.
HAO Yuan-fang, WU Ying-jie, HE Lian. THE METHOD RESEARCH TO DETERMINE ANIONS IN ENVIRONMENTAL WATER SAMPLES WITH ION CHROMATOGRAPHY. GEOLOGY AND RESOURCES, 2005, 14(2): 149-150,148.