首页 | 本学科首页   官方微博 | 高级检索  
     检索      

PANDOROS高压发生器Jmax灯亮故障检修
引用本文:李国耀,张延武.PANDOROS高压发生器Jmax灯亮故障检修[J].CT理论与应用研究,1995,4(4):34-35.
作者姓名:李国耀  张延武
作者单位:广州红十字会医院,广州军医总医院
摘    要:本文分析了引发Jmax灯亮的三种原因。即X线管电流检测回路R33电位计故障;KV参数值产生回路V24管击穿故障,以及灯丝初级加热回路V2管软击穿故障,所以导致的高压发生器低电流保护,Jmax灯亮,不同的是前二项原因使曝光开始即刻Jmax灯亮。

关 键 词:低电流保护  软击穿  高压发生器  Jmax灯亮  故障

Trouble-Shooting of "J_(max)" lighted in PANDOROS High Voltage Generator
Li Guoyue,Zhang Yanwu Guang-zhou Red Cross Hospital Guang-zhou Liu Hua Hospital.Trouble-Shooting of "J_(max)" lighted in PANDOROS High Voltage Generator[J].Computerized Tomography Theory and Applications,1995,4(4):34-35.
Authors:Li Guoyue  Zhang Yanwu Guang-zhou Red Cross Hospital Guang-zhou Liu Hua Hospital
Institution:Li Guoyue;Zhang Yanwu1) Guang-zhou Red Cross Hospital 510220) 2) Guang-zhou Liu Hua Hospital 510010
Abstract:Three causes that can make "Jmax" bight were analysed in the Paper. They arefault of the Potentiometer (R33) in the current detection circuitry of X-ray. the breakdown of the V24 in circuit of KV reference PCB, and the soft-breakdown of the V2 in the heating circuitof prdriary of Heater. All this may causes the low current protection. But the three reasons arediferent. The "Jmax" is lighted the instant when exposure for first and second reason, and during the exposure for the third.
Keywords:s: Low current protection  Soft-breakdown  High voltage generator