Nolan–Pollak type CN counters in the Vienna aerosol workshop
Authors:J L Gras  J Podzimek  T C O'Connor  K -H Enderle
Abstract:Three “standard” Nolan–Pollak (N–P) and a modified N–P design condensation nucleus (CN) counters were included in the Vienna Workshop on Intercomparison of Condensation Nuclei and Aerosol Particle counters. These counters came from diverse backgrounds, namely programs in USA, Europe and Australia. In this work, principles of the operation and previous history of calibration of the N–P expansion counter are briefly reviewed and comparisons between the particular counters used in the workshop are presented and discussed. Counting agreement was found to be very good between the N–P counters, typically better than ±12% for a range of aerosol sizes and compositions from a minimum diameter of 4 nm. The independently calibrated GIV CNC-440 (modified N–P type counter) also agreed well with the N–P counters. The minimum size sensitivity of the N–P counter was examined showing a lower detection limit for insoluble (Ag) particles of around 2.6±0.3 nm diameter.
Keywords:Aerosol number concentration  Detection efficiency  Condensation nuclei counter  Intercomparison study
本文献已被 ScienceDirect 等数据库收录!