Determination of the distribution of size of irregularly shaped particles from laser diffractometer measurements
Authors:Claudio L Schneider  Reiner Neumann  Alan S Souza
Institution:1. CETEM, Brazil;2. Companhia Vale do Rio Doce, Brazil
Abstract:The formal stereological transformation equation for particle sieve size distribution from measurements in lower dimensional spaces is applied to laser diffractometer measurements. The transformation function for iron ore particles is measured experimentally, and modeled. The solution is tested against the measured transformation function data as well as synthetic composite distributions of the original sample. The natural size distribution of a sample taken from a grinding circuit stream was measured by a combination of standard sieving and cyclosizer, and the result is compared to the transformed size distribution calculated from laser diffractometer measurements. The stereological transformation technique performed well in all cases.
Keywords:Stereological transformation  Sub-sieve size measurement  Laser diffractometer