Influence of the Sonic Anemometer Temperature Calibration on Turbulent Heat-Flux Measurements
Authors:Renzo Richiardone  Massimiliano Manfrin  Silvia Ferrarese  Caterina Francone  Vito Fernicola  Roberto Maria Gavioso  Luca Mortarini
Institution:1.Dipartimento di Fisica,Università di Torino,Torino,Italy;2.Dipartimento di Ingegneria Aerospaziale,Politecnico di Torino,Torino,Italy;3.INRIM—Istituto Nazionale di Ricerca Metrologica,Torino,Italy;4.ISAC-TO Istituto di Scienze dell’Atmosfera e del Clima—Sezione di Torino,Torino,Italy
Abstract:The speed of sound in moist air is discussed and a more accurate value for the coefficient of the linear dependence of sonic temperature on specific humidity is proposed. An analysis of speed-of-sound data measured by three sonic anemometers in a climate chamber and in the field shows that the temperature response of each instrument significantly influences not only the determination of sonic temperature, but also its fluctuations. The corresponding relative contribution to the error in the evaluation of the temperature fluctuations and the turbulent heat fluxes can be as high as 40%. The calibration procedure is discussed and a method of correction is proposed.