山西定襄地震台FH-463A与BHC-336定标器测氡仪系统观测数据对比
引用本文:刘俊芳,范雪芳,郭宝仁,刘金柱,郭宇,高文玉.山西定襄地震台FH-463A与BHC-336定标器测氡仪系统观测数据对比[J].地震地磁观测与研究,2021,42(3):214-224.
作者姓名:刘俊芳  范雪芳  郭宝仁  刘金柱  郭宇  高文玉
作者单位:1.中国太原 030021 山西省地震局;2.中国山西 034200 忻州地震监测中心站
基金项目:中国地震局监测、预报、科研三结合课题(项目编号:3JH-201901076)
摘    要:对山西定襄地震台2套FD-125氡钍分析器及其配套定标器分别为FH-463A与BHC-336的测氡仪系统,从定标器阈值设定、观测数据稳定性及数据一致性方面,开展定标器阈值设定对比试验和水氡测值同步对比观测。结果表明:FH-463A定标器测氡仪系统水氡数据标准误差更小,精度更高,稳定性更佳;BHC-336定标器阈值为0.7 V的测氡仪系统,其主样和副样的标准偏差和均值标准误差最小,较其他阈值观测数据稳定;BHC-336定标器与FH-463A定标器分别使用0.7 V阈值和2.0 V阈值,其测氡仪系统水氡观测数据的同步性和稳定性较好。按2014年修订的学科规范,确定BHC-336定标器选择0.7 V阈值时可获取稳定的观测数据。

关 键 词:测氡仪  定标器  阈值  对比观测  统计学

Comparison of observations by two radon measuring systems with FH-463A and BHC-336 calibrators at Dingxiang Seismic Station of Shanxi Province
LIU Junfang,FAN Xuefang,GUO Baoren,LIU Jinzhu,GUO Yu,GAO Wenyu.Comparison of observations by two radon measuring systems with FH-463A and BHC-336 calibrators at Dingxiang Seismic Station of Shanxi Province[J].Seismological and Geomagnetic Observation and Research,2021,42(3):214-224.
Authors:LIU Junfang  FAN Xuefang  GUO Baoren  LIU Jinzhu  GUO Yu  GAO Wenyu
Institution:1.Shanxi Earthquake Agency, Taiyuan 030021, China;2.Xinzhou Earthquake Monitoring Center Station, Shanxi Province 034200, China
Abstract:For two sets of FD-125 radon thorium analyzer at Dingxiang Seismic Station of Shanxi Province, whose matching calibrators are FH-463A and BHC-336 respectively, this paper carried out a comparison test of scaler threshold setting and synchronous observation of water radon from the aspects of scaler threshold setting, observation data stability, and data consistency. The results show that the standard error of water radon observation by the radon measuring instrument with the FH-463A calibrator is smaller, the accuracy is higher, and the stability is better. The standard deviation and mean standard error of the primary and secondary samples observed by the radon measuring instrument with the BHC-336 calibrator are the smallest when the threshold is 0.7 V. The observation is more stable than other thresholds. The water radon observations have better synchronization and stability when the radon measuring instrument with the BHC-336 calibrator uses a 0.7 V threshold and the radon measuring instrument with the FH-463A calibrator uses a 2.0 V threshold. According to the discipline specification revised in 2014, it is determined that stable observation data can be obtained when the 0.7 V threshold value is selected for the BHC-336 calibrator.
Keywords:radon measuring instrument  scaler  threshold  comparative observation  statistical