首页 | 本学科首页   官方微博 | 高级检索  
     检索      


QUANTITATIVE ANALYSIS OF TRACE ELEMENTS IN GEOLOGICAL MATERIALS BY LASER ABLATION ICPMS: INSTRUMENTAL OPERATING CONDITIONS AND CALIBRATION VALUES OF NIST GLASSES
Authors:MD NORMAN  NJ PEARSON  A SHARMA  WL GRIFFIN
Institution:Key Centre for Geochemical Evolution and Metallogeny of Continents (GEMOC), School of Earth Sciences, Macquarie University, Sydney 2109, Australia;CSIRO Exploration and Mining, P.O. Box 136, North Ryde 2113, Australia
Abstract:A UV laser ablation microprobe coupled to an ICPMS has been used to determine trace element concentrations in solids with a spatial resolution of 50 microns and detection limits ranging from 2 μg/g for Ni to 50 ng/g for the REE, The, and U. Experiments designed to optimize laser operating conditions show that pulse rates of 4 Hz produce a steady state signal with less inter-element fractionation per unit time than higher pulse rates (10–20 Hz). Comparisons of laser microprobe analyses of garnets and pyroxenes using the NIST 610 and 612 glasses as calibration standards, with proton microprobe, solution ICPMS, INAA and XRF data show no significant matrix effects. Laser microprobe analyses of the NIST 610 and 612 glasses have a precision and accuracy of 2–5%, and error analysis shows that counting statistics and the precision on the internal standard concentration accounts for the analytical uncertainty. The NIST glasses appear to be useful calibration materials for trace element analysis of geological materials by laser microprobe.
Keywords:
设为首页 | 免责声明 | 关于勤云 | 加入收藏