首页 | 本学科首页   官方微博 | 高级检索  
     检索      


The Sensitivity of Four Monitoring Well Sampling Systems to Low Concentrations of Three Volatile Organics
Authors:Ronald Schalla  David A Myers  Mary Ann Simmons  John M Thomas  Antony P Toste
Abstract:Four state-of-the-art ground water sampling systems were analyzed to determine their reliability in providing representative samples of the volatile chlorinated hydrocarbons trichloroethylene (TCE), perchloroethylene (PCE), and 1,1,1-trichloroethane (TCA) from a simulated monitoring well. The sampling systems studied represent four commonly used devices, including a stainless steel and Teflon® piston pump, a Teflon bailer, a Teflon bladder pump, and a PVC air-lift pump.
Controlled laboratory sampling experiments were conducted in a tank and well test chamber designed to approximate field conditions. A well purging and sampling procedure was used in the test apparatus to determine the accuracy and precision of each device for detecting low concentrations of the compounds in ground water. The compounds selected are some of the most ubiquitous hazardous contaminants found in shallow aquifers near hazardous waste sites throughout the United States.
No significant statistical difference was found among the four sampling systems in detecting the compounds.
Keywords:
设为首页 | 免责声明 | 关于勤云 | 加入收藏