首页 | 本学科首页   官方微博 | 高级检索  
     检索      


MKIIIB-CTD: improving its system output
Authors:Thomas J Mller  Jürgen Holfort  Frank Delahoyde  Robert Williams
Institution:Thomas J. Möller, Jürgen Holfort, Frank Delahoyde,Robert Williams
Abstract:In the Neil Brown Instruments' MKIIIB-CTD (conductivity-temperature-depth profiler), the system's digital outputs for the three basic measurements of temperature, conductivity and pressure typically show some small amplitude deviations from smooth calibrations which should be corrected for to achieve high accuracies, as required, e.g. within the Hydrographic Program (WHP) of the current World Ocean Circulation Experiment (WOCE). These deviations show up as (i) a strong nonlinearity or even discontinuity of several mK close to 0°C in temperature output leading to too high subzero temperatures; (ii) a jump of order 0.002 mS cm−1 in conductivity output when passing the half-range value 32.768 mS cm−1, which causes jumps in the relation of potential temperature and salinity; and (iii) errors in pressure measurements of up to 4 dbar due to mechanical hysteresis and both static and dynamic responses to temperature changes. The existence of these effects is demonstrated, and methods to reduce the associated errors are suggested.
Keywords:
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏