首页 | 本学科首页   官方微博 | 高级检索  
     检索      

基于X射线连续谱的光子计数探测器能量标定方法
引用本文:李中浩,贠明凯,姜小盼,张志都,程志威,魏存峰,史戎坚,魏龙,王哲.基于X射线连续谱的光子计数探测器能量标定方法[J].CT理论与应用研究,2018,27(3):363-372.
作者姓名:李中浩  贠明凯  姜小盼  张志都  程志威  魏存峰  史戎坚  魏龙  王哲
作者单位:1. 中国科学院高能物理研究所, 北京市射线成像技术与装备工程技术研究中心, 北京 100049;
基金项目:国家重点研发计划(2016YFC0100400),中国科学院科研装备研制项目(YZ201511),中国科学院关键技术研发团队项目(GJJSTD20170005)
摘    要:随着探测器技术的发展,具有能量分辨能力的光子计数探测器在X射线能谱CT成像中应用越来越广泛。能量与阈值的对应关系对能谱CT成像具有重要意义,受到实验条件的限制,目前的光子计数探测器的能量标定方法有一定的局限性。针对光子计数探测器的单像素标定问题,本文提出一种基于X射线光机连续谱的能量标定方法,通过建立和分析X光谱模型,利用光机连续谱中高能端的能量信息,得到阈值与能量之间的关系。实验结果表明该方法操作方便,对探测器单像素标定结果较为准确,对其他类型光子计数探测器的标定具有参考意义。 

关 键 词:X射线光机连续谱    光子计数探测器    能量标定
收稿时间:2018-02-26

Energy Calibration Method of the Photon Counting Detector Based on Continuous X-ray Spectrum
LI Zhong-hao,YUN Ming-kai,JIANG Xiao-pan,ZHANG Zhi-du,CHENG Zhi-wei,WEI Cun-feng,SHI Rong-jian,WEI Long,WANG Zhe.Energy Calibration Method of the Photon Counting Detector Based on Continuous X-ray Spectrum[J].Computerized Tomography Theory and Applications,2018,27(3):363-372.
Authors:LI Zhong-hao  YUN Ming-kai  JIANG Xiao-pan  ZHANG Zhi-du  CHENG Zhi-wei  WEI Cun-feng  SHI Rong-jian  WEI Long  WANG Zhe
Institution:1. Beijing Engineering Research Center of Radiographic Techniques and Equipment, Institute of High Energy Physics, Chinese Academy of Sciences, Beijing 100049, China;2. School of Nuclear Science and Technology, University of Chinese Academy of Sciences, Beijing 100049, China
Abstract:With the development of the detector technology, the energy discriminating photon counting detectors have been widely used in Spectral CT imaging. The relationship between energy and threshold is very important for Spectral CT Imaging. Being subjected to experiment conditions, the current energy calibration methods of photon counting detector have some limitations. In order to solve single pixel energy calibration of photon counting detector, this paper describes an energy calibration methods using continuous X-ray Spectrum. Energy information of the high energy end in continuous spectrum is used to determine the relationship between energy and threshold by building and analyzing X-ray model. Experimental results show that this method has the advantages of convenient operation and relatively accurate result, which can also be referenced by other energy calibration methods. 
Keywords:
本文献已被 万方数据 等数据库收录!
点击此处可从《CT理论与应用研究》浏览原始摘要信息
点击此处可从《CT理论与应用研究》下载免费的PDF全文
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号