Stability of Temperature and Conductivity Sensors of Argo Profiling Floats
Authors:Eitarou Oka  Kentaro Ando
Institution:(1) Frontier Observational Research System for Global Change, Natsushima-cho, Yokosuka, Kanagawa 237-0061, Japan;(2) Japan Marine Science and Technology Center, Natsushima-cho, Yokosuka, Kanagawa 237-0061, Japan
Abstract:After recalibration of the temperature and conductivity sensors of three Argo profiling floats recovered after operations for four to nine months, the results indicate that the floats basically showed no significant drift, either in temperature or salinity, and adequately fulfilled the accuracy requirement of the Argo project (0.005°C for temperature and 0.01 psu for salinity). Only the third float showed a significant offset in salinity of about −0.02 psu, as expected from comparison between the float data and the shipboard conductivity-temperature-depth data. This offset was caused by the operational error of the PROVOR-type float, in which the surface water was pumped immediately after the launch, fouling the conductivity sensor cell. This revised version was published online in July 2006 with corrections to the Cover Date.